Sodium hypochlorite (NaClO) is the most commonly used high-efficiency oxidant in drinking water disinfection, swimming pool water, hospital wastewater, industrial circulating water and bleaching processes. In actual dosing control, ORP (oxidation-reduction potential) is the core parameter that indirectly reflects the residual chlorine level and disinfection effectiveness — after sodium hypochlorite is dosed, the water's ORP rises; the more stable the ORP, the more controllable the disinfection. However, sodium hypochlorite itself is a strongly oxidizing medium that continuously erodes the platinum sensing membrane and reference system of ordinary ORP electrodes; long-term operation can easily lead to potential drift, slower response, and even electrode poisoning and failure.
For sodium hypochlorite solution conditions, Chunye Technology has launched the CS2505C/CS2505CT glass-body ORP electrodes and the CS2705FC/CS2705FCT plastic-body ORP electrodes, paired with the T3000 online ORP meter, forming a complete online ORP monitoring solution for sodium hypochlorite water quality.
1. Sodium Hypochlorite Water ORP Measurement: Why Do Ordinary Electrodes Easily "Crack Under Pressure"?

The strong oxidizing nature of sodium hypochlorite solution imposes four special requirements on ORP electrodes:
Platinum sensing membrane resistant to chloride attack: Sodium hypochlorite continuously releases active chlorine. After prolonged contact, ordinary platinum membranes are prone to surface oxidation passivation, reducing the electron exchange rate and leading to sluggish response and low readings. The Chunye CS2505/CS2705 series adopts a platinum (Pt) sensing membrane that maintains stable electron exchange within the ±1000 mV range, adapting to the sodium hypochlorite oxidizing environment.
Oxidation-resistant reference system: Active chlorine can penetrate along the liquid junction into the reference chamber, contaminating the reference electrolyte and causing potential drift. Dedicated sodium hypochlorite ORP electrodes feature targeted design in the liquid junction structure and reference formulation to slow down chlorine contamination.
Temperature fluctuation compensation: ORP is temperature-sensitive, and sodium hypochlorite disinfection processes vary greatly with seasons and operating conditions. The CT models feature a built-in temperature sensor (NTC10K/NTC2.2K/PT100/PT1000) that can be directly connected to the T3000 for automatic temperature compensation.
Long-term stability and interlocked control: Disinfection dosing requires continuous online monitoring with support for over-limit alarms and dosing pump interlocking, which requires the electrode and instrument to be fully matched in signal stability and output interfaces.
2. Chunye Dedicated Sodium Hypochlorite ORP Electrodes: Two Body Types, One Set of Technical Standards

The CS2505 and CS2705 series share the same set of core technical parameters, differing only in body material, thread and cable length:
表格
Sensing membrane | Pt (platinum) | Pt (platinum) |
ORP range | ±1000 mV | ±1000 mV |
Temperature range | 0–80 °C | 0–80 °C |
Pressure rating | 0–0.6 MPa | 0–0.6 MPa |
Body | Glass | PP+GF |
Connection thread | PG13.5 | NPT 3/4 (bottom), G 3/4 (top) |
Cable length | 5 m or as agreed | 10 m or as agreed |
Temperature sensor (CT models) | NTC10K/NTC2.2K/PT100/PT1000 | NTC10K/NTC2.2K/PT100/PT1000 |
Selection highlights:
Glass-body CS2505C/CS2505CT: The corrosion-resistant glass body and PG13.5 thread fit mainstream industrial protective housings, suitable for pressurized installation scenarios such as pipelines and reaction tanks;
Plastic-body CS2705FC/CS2705FCT: The PP+GF plastic body is impact-resistant and durable, with dual threads of NPT 3/4 (bottom) / G 3/4 (top), suitable for on-site installation in open water basins, dosing tanks, etc.;
CT models (CS2505CT/CS2705FCT): Feature a built-in temperature sensor; automatic temperature compensation is enabled when connected to the T3000 — the first choice for process sections with large temperature fluctuations. C models have no temperature compensation, suitable for stable-temperature, cost-sensitive scenarios.
3. T3000 Online ORP Meter: Stable ORP Signal Output with Direct Dosing Interlock

The electrode requires a matching online transmitter. The T3000 online ORP meter is a microprocessor-based industrial water quality monitoring and control instrument, configurable with ORP electrodes and widely used in power plants, petrochemical, environmental water treatment, food and beverage, aquaculture and other industries.
Measurement performance:
ORP range −2000 to +2000 mV, resolution 1 mV, basic error ±2 mV, stability ≤2 mV/24 h, temperature −10 to 150.0 °C;
Differential signal measurement mode, resistant to on-site interference, stable and reliable in long-term operation.
Control and output:
Two groups of relay control switches, supporting high-limit, low-limit and hysteresis control, can directly interlock the sodium hypochlorite dosing pump and the over-limit alarm;
Current output: 1 channel standard, optional 2 channels of 0/4–20 mA (load resistance <750 Ω), with RS485 MODBUS RTU communication for connection to PLC/SCADA;
Dot-matrix LCD monochrome display, smart menu operation, multiple automatic calibration modes, manual/automatic temperature compensation, simultaneous display of ORP, temperature and status on one screen, password protection against accidental operation.
Installation and power supply:
Dimensions 98 × 98 × 130 mm, panel-mount/wall-mount installation, panel cutout 92.5 × 92.5 mm;
Power supply optional 85–265 VAC ±10% or 9–36 VDC, power ≤3 W;
Protection rating IP65, unit weight 0.6 kg, suitable for humid dosing room environments.
4. Selection Recommendations and Ordering Codes

Three steps:
Choose the body by installation position: for pressurized scenarios such as pipelines and reaction tanks, choose the glass-body CS2505CT (PG13.5); for on-site scenarios such as open water basins and dosing tanks, choose the plastic-body CS2705FCT (NPT 3/4 / G 3/4);
Decide on temperature compensation: for large temperature fluctuations, prioritize the CT models, which enable automatic compensation when connected to the T3000;
Choose the ordering code: combine three segments of "temperature sensor + cable length + cable connector", e.g., CS2505CT-N1-m10-A4 means a CS2505CT electrode, NTC10K temperature sensor, 10 m cable, BNC connector.
表格
Temperature sensor | N0 none / N1 NTC10K / N2 NTC2.2K / P1 PT100 / P2 PT1000 |
Cable length | m5 (5 m) / m10 (10 m) / m15 (15 m) / m20 (20 m) |
Cable connector | A1 tinned wire end / A2 Y spade terminal / A3 flat pin / A4 BNC |
5. Frequently Asked Questions (FAQ)

Q1: Why must dedicated ORP electrodes be used in sodium hypochlorite environments? Can ordinary pH electrodes be used?A: No, they cannot be used interchangeably. Sodium hypochlorite is a strong oxidant; the platinum membrane of ordinary electrodes is prone to passivation, and the reference is easily contaminated by chlorine. The Chunye CS2505/CS2705 series are dedicated ORP electrodes designed for sodium hypochlorite solutions, with the platinum sensing membrane and liquid junction structure both adapted for chloride resistance.
Q2: What is the difference between CS2505C and CS2505CT, and between CS2705FC and CS2705FCT?A: The C models have no temperature sensor, while the CT models feature a built-in NTC10K/NTC2.2K/PT100/PT1000 temperature sensor that can be directly connected to the T3000 to enable automatic temperature compensation.
Q3: What ORP value indicates adequate sodium hypochlorite disinfection?A: The thresholds vary by process. Generally, ORP ≥ 650 mV for drinking water disinfection and ≥ 700 mV for swimming pool water can serve as reference ranges; specifics should be based on on-site process calibration. The T3000 supports setting high limits, low limits and hysteresis according to actual site conditions to control the dosing pump.
Q4: How often should the electrode be calibrated?A: In sodium hypochlorite environments, monthly calibration with ORP standard solutions (e.g., 220 mV or 465 mV standard solutions) is recommended. If the platinum membrane is passivated, briefly immerse it in 0.1 mol/L dilute hydrochloric acid and then rinse with pure water.
Disclaimer: This article is a technical introduction for reference. Specific electrode selection is subject to on-site water quality, temperature and installation conditions; product parameters are subject to the latest information from Chunye Technology.